Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary : the device was received for evaluation.Visual inspection under magnification doesn¿t show any chips or burrs in the tool that would indicate that metal shavings have come from the device.The device looks worn but is in as expected condition.This complaint cannot be confirmed no further information regarding the cause of the defect has been provided to help determine a root cause for this failure.Also, no lot number was provided to depuy mitek, therefore no device history record review could be performed.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).
|