• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN MFG DC BOULDER COOL-TIP; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MFG DC BOULDER COOL-TIP; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number ACT2020
Medical Device Problem Code Peeled/Delaminated (1454)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 09/17/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, post-operatively, the coating part was found damaged, but nothing disengaged from the device.Reoperation was not required.There was no patient involved.
 
Additional Manufacturer Narrative
Evaluation summary: one device was received for evaluation.The returned product did not meet specification as received.Visual inspection found the clear insulation was damaged.The reported condition was confirmed.The investigation of the device found damaged insulation on the needle.The hole was most likely caused by the needle contacting the guiding needle.The device failed hipot testing.The investigation identified the root cause of the reported event to be the needle contacting the metallic guiding needle.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOL-TIP
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
MDR Report Key9170264
Report Number1717344-2019-01289
Device Sequence Number16848734
Product Code GEI
UDI-Device Identifier10884524001210
UDI-Public10884524001210
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K973297
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberACT2020
Device Catalogue NumberACT2020
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/16/2019
Initial Date Received by Manufacturer 09/17/2019
Supplement Date Received by Manufacturer10/28/2019
Initial Report FDA Received Date10/09/2019
Supplement Report FDA Received Date11/12/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-