The manufacturer did not receive x-rays, or other source documents for review.The manufacturer did not receive the device for investigation.As no lot number was provided, the device history records could not be reviewed.Attempts to obtain additional information have been made; however, no more is available.A cause for this specific event cannot be ascertained from the information provided.As soon as supplemental information becomes available, an updated report will be submitted.Zimmer biomet¿s reference number of this file is (b)(4).Lot number unknown.
|
|
Investigation results were made available.Additional: h2, h6.Correction: b4, b5, g4, g7, h10.Dhr-review: as no lot number was provided, the device history records could not be reviewed.The missing device information has been requested but was not available.At zimmer gmbh all medical devices prior release to market undergo several quality inspections as defined in our quality procedures.Our quality inspection- and deviation procedures ensure that only products fulfilling the specification are sold.These procedures are part of the overall quality management system at zimmer gmbh and get regularly audited by our notified body, competent authorities and internal and external auditors.Thus, for all products sold to the market can be assumed having a complete and correct dhr.Trend analysis: no trend has been identified.Event description: it was reported that the patient was implanted with a ncb plate on unknown date and revised due to implant fracture on unknown date.Review of received data: - no medical data such as x-rays, surgical notes or any other case-relevant documents received.Devices analysis: - no product was returned to zimmer biomet for in-depth analysis.Review of product documentation: - this device is intended for treatment.Conclusion summary: it was reported that the patient was implanted with a ncb plate on unknown date and revised due to implant fracture on unknown date.The dhr check could not be performed as the lot number was not available.Neither x-rays, operative notes, office visit notes, nor devices or photos of the explanted implant were received; therefore the condition of the component is unknown.Patient factors that may affect the performance of the components such as bone quality, activity level, type of activity (low impact vs.High impact), and relevant medical history are unknown.Adherence to rehabilitation protocol is unknown.The investigation results did not identify a non-conformance or a complaint out of box (coob).Based on the given information and the results of the investigation, the complaint could not be confirmed as the alleged failure could not be identified or reproduced.The need for corrective measures is not indicated and zimmer gmbh considers this case as closed.Zimmer biomet's reference number of this file is (b)(4).
|