| Model Number |
MMT-523LNAB |
| Medical Device Problem Code |
Device Difficult to Program or Calibrate (1496)
|
| Health Effect - Clinical Code |
Complaint, Ill-Defined (2331)
|
| Date of Event |
07/31/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Customer reported via phone call that the insulin pump act button take forever to push.The customer¿s blood glucose level was unknown at the time of the incident.Troubleshooting was performed.Customer was advised that the insulin pump needed to be replaced and was advised to discontinue use of the insulin pump and to revert to the backup plan per health care professional instructions.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Insulin pump received with intermittent button response due to corroded keypad traces.No button error alarm during testing.Device passed self test.
|
| |
|