| Model Number |
MMT-723RNAS |
| Medical Device Problem Code |
Priming Problem (4040)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/02/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer's wife reported via phone call that his insulin pump was stuck on fill tubing and unable to exit the preparing to prime loop.Blood glucose was unknown.Troubleshooting was performed.Customer stated they did not receive 2nd series of beeps nor did numbers appear on the screen.The customer was advised that the insulin pump needed to be replaced and was advised to discontinue use of the insulin pump and to revert to the backup plan.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed rewind test, basic occlusion test, occlusion test, prime/a33 test, excessive no delivery test and displacement test.No unexpected prime/fill anomaly noted during testing.The test reservoir locked in place properly and no anomaly noted.(b)(4).
|
| |
|