Model Number 71342340
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Injury (2348)
Date of Event 12/19/2018
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported recurrent left hip dislocation and greater trochanter fracture on (b)(6) 2018 and (b)(6) 2018.Open reduction performed to correct the problem.
 
Additional Manufacturer Narrative
The devices, used in treatment were not returned for evaluation, reporting event could not be confirmed.A clinical evaluation was conducted and confirms patient sustained a greater trochanter fracture intraoperatively, during his fourth revision.It cannot be concluded that the reported open reduction internal fixations were associated with a mal-performance of the implant vs early ambulation a complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.Factors and/or potential probable causes that could have contributed to the reported event include but are not limited to incorrect material specification, traumatic injury, and/or patient condition.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further actions are being taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
40MM OXINIUM MODULAR HEAD
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key9171678
Report Number1020279-2019-03604
Device Sequence Number11368839
Product Code MBL
UDI-Device Identifier00885556112410
UDI-Public00885556112410
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number71342340
Device Catalogue Number71342340
Initial Date Received by Manufacturer 09/10/2019
Supplement Date Received by Manufacturer09/10/2019
Initial Report FDA Received Date10/09/2019
Supplement Report FDA Received Date04/14/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-