Model Number 2426-0500 |
Device Problems
Fluid/Blood Leak (1250); Tear, Rip or Hole in Device Packaging (2385); Deformation Due to Compressive Stress (2889)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The affected product has been received and the evaluation is pending.A follow up report will be submitted once the evaluation is completed.
|
|
Event Description
|
It was reported that the nurse was attempting to prime the tubing set when the tubing set was noted to be leaking at the safety clamp.On further review of the tubing set, it was found that there was a kink at the safety clamp that created a hole in the tubing set.The customer further stated that there was no patient involvement.
|
|
Event Description
|
It was reported that the rn was attempting to prime the tubing set when the tubing set was noted to be leaking at the safety clamp.On further review of the tubing set, it was found that there was a kink at the safety clamp that created a hole in the tubing set.The customer further stated that there was no patient involvement.
|
|
Manufacturer Narrative
|
The customer¿s report that the tubing set leaked and was kinked at the safety clamp was confirmed.Visual inspection confirmed a kink on the tubing at the engagement of the lower fitment¿s outlet port.Closer inspection of the kink observed that the tubing was incorrectly inserted into the lower fitment¿s outlet port, with the tubing¿s fluid path opened and not attached at the engagement.No visible damages, holes, or other issues were observed.Functional testing could not be performed due to the separation and the obvious leaking that would occur.The root cause that the tubing set leaked and was kinked at the safety clamp is due to equipment/operator disassembly error that occurred during the manufacturing process.
|
|
Search Alerts/Recalls
|