ANGIOMED GMBH & CO. MEDIZINTECHNIK KG COVERA VASCULAR COVERED STENT SYSTEM Back to Search Results
Model Number AVFM08060
Medical Device Problem Codes Obstruction of Flow (2423); Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code Thrombosis (2100)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Medical images were provided to the manufacturer.The lot number for the device was provided.The device history records are currently under review.The device is not available for return.The investigation is currently underway.
 
Event or Problem Description
It was reported that during a covered stent placement procedure within the cephalic arch, the patient allegedly experienced thrombosis after post dilating the covered stent.It was further reported that the clot was removed from within the stent.The patient was reportedly stable post clot removal.
 
Additional Manufacturer Narrative
H10: manufacturing review: the lot history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the physical sample was not available; x-ray images were provided documenting the placed covered stent including flow obstruction.However, an indication for an irregular placement or a defect of the placed stent could not be found.An indication for a device relation of the flow obstruction could not be determined.Therefore, the investigation of this issue is closed with inconclusive result.Based on the information available a definite root cause for the event reported could not be determined.Labeling review: in reviewing the relevant labeling for this product, it was found that the instructions for use (ifu) sufficiently address the potential issue.Regarding pre- and post- dilation, the ifu state: 'pre-dilate the stenosis with a pta balloon catheter of appropriate length and diameter for the lesion to be treated' and 'post dilate the covered stent with an angioplasty balloon sized appropriately as to ensure complete wall apposition to the reference vessel.' the ifu further state 'potential complications may include, but are not limited to: thrombotic occlusion, restenosis requiring reintervention (.)'.Holding and handling of the system throughout deployment was found sufficiently described.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during a covered stent placement procedure within the cephalic arch, the patient allegedly experienced thrombosis after post dilating the covered stent.It was further reported that the clot was removed from within the stent.The patient was reportedly stable post clot removal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COVERA VASCULAR COVERED STENT SYSTEM
Common Device Name
VASCULAR COVERED STENT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstrasse 6
karlsruhe 76227
GM  76227
MDR Report Key9175423
Report Number9681442-2019-00157
Device Sequence Number17040817
Product Code PFV
UDI-Device Identifier00801741106514
UDI-Public(01)00801741106514
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P170042
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 12/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/18/2021
Device Model NumberAVFM08060
Device Catalogue NumberAVFM08060
Device Lot NumberANDN1872
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/13/2019
Supplement Date Received by Manufacturer11/29/2019
Initial Report FDA Received Date10/10/2019
Supplement Report FDA Received Date12/03/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-