Catalog Number 8065752201
Device Problem Material Split, Cut or Torn (4008)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/04/2019
Event Type  malfunction  
Manufacturer Narrative
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A physician reported that the tip of the irrigation/aspiration sleeve was found to be torn during a procedure.The product was replaced and the procedure was completed.There was no patient harm.
 
Manufacturer Narrative
The lot complaint history was reviewed, this is the third complaint for the finish goods lot; however, the first for this issue for this lot.The device history record shows the product was released per specifications.The customer returned two infusion sleeves.The first infusion sleeve with was visually inspected and it was found that the tip of the sleeve was completely torn off.No damage was identified with the second infusion sleeve.The torn off tip shaft was not returned.Microscopic examination found the damage is consistent with damage that occurs when a phaco tip snags the sleeve during mounting of the sleeve onto the phaco tip.The root cause of the customer¿s complaint is related to user error.The entire tip and shaft of the infusion sleeve was torn off which is consistent with damaged observed when the sleeve is mounted improperly during setup of the handpiece.No action will be taken for this occurrence, as the root cause is related to user error of the product.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION SURGICAL PROCEDURE PAK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
MDR Report Key9175536
MDR Text Key182968622
Report Number1644019-2019-00152
Device Sequence Number1
Product Code LRO
Combination Product (y/n)N
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 12/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/10/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2021
Device Catalogue Number8065752201
Device Lot Number2251134H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/15/2019
Date Manufacturer Received12/19/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
CENTURION VISION SYSTEM
-
-