MAUDE Adverse Event Report: ROCHE DIABETES CARE, INC. ACCU-CHEK ® AVIVA; BLOOD GLUCOSE MONITORING DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIABETES CARE, INC. ACCU-CHEK ® AVIVA; BLOOD GLUCOSE MONITORING DEVICE Back to Search Results
Catalog Number 06870287001
Medical Device Problem Code Incorrect Measurement (1383)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/16/2019
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the meter reads in mmol/l rather than mg/dl.Results in the meter memory indicates that readings are stored as mg/dl.
 
Additional Manufacturer Narrative
This supplemental mdr is being submitted as a part of a retrospective review and remediation effort based on errors that occurred in the submission of mdrs for incorrect manufacturer name and/or address.A non-conformance report (ncr) ¿ 16847 to implement corrective actions was opened on january 29, 2025.Device performance and/or risk profile are not impacted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCU-CHEK ® AVIVA
Common Device Name
BLOOD GLUCOSE MONITORING DEVICE
Manufacturer (Section D)
ROCHE DIABETES CARE, INC.
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
SANMINA-SCI INDIA PVT LTD.
oz-1 sipcot hi-tech sez,oragad
sriperumpudur taluk
chennai, tamil nadu 60210 5
IN   602105
Manufacturer Contact
greg smith
9115 hague road
na
indianapolis, IN 46250-0457
3175212484
MDR Report Key9175798
Report Number3011393376-2019-03769
Device Sequence Number11733615
Product Code NBW
UDI-Device Identifier00365702544109
UDI-Public00365702544109
Combination Product (Y/N)N
Initial Reporter StateAS
Initial Reporter CountryUS
PMA/510(K) Number
K043474
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/23/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number06870287001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/09/2019
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 09/16/2019
Supplement Date Received by Manufacturer12/02/2019
07/22/2025
Initial Report FDA Received Date10/10/2019
Supplement Report FDA Received Date12/18/2019
07/23/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Patient Age64 YR
Patient SexFemale
-
-