| Model Number |
27026EF |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/13/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product was evaluated and confirmed the distal deflector mechanism is broken off and there is corrosion around the area.The date code of the product is 2015.The instrument has been in use for 4 years.The tip was retrieved from inside the patient and the procedure completed with no harm to the patient.
|
| |
|
Event or Problem Description
|
|
Allegedly per the customer, during a cystoscopy with biopsy the tip of the deflector mechanism broke off while inside the patient.The tip was retrieved in one piece with no harm to the patient.Once the piece was retrieved the procedure continued and completed.
|
| |
|
Additional Manufacturer Narrative
|
|
The correct product number is 2726ef.
|
| |
|