| Catalog Number |
56535RIM |
| Medical Device Problem Codes |
Fracture (1260); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
09/16/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The suspect device was not returned for evaluation however a photo was provided and reviewed.The complaint was confirmed.The root cause was undetermined however the photo shows evidence that the tip may have split due to degradation consistent with some type of chemical or environmental exposure.A search of the complaint database was performed and no similar complaints for this lot number were found.The device history record was reviewed and no exception documents were found.
|
| |
|
Event or Problem Description
|
|
The account alleges that during a percutaneous transluminal angioplasty procedure the catheter tip broke into small pieces.No additional procedures/interventions were needed for this patient.No patient injury to report.
|
| |
|