ZOLL MEDICAL CORPORATION M SERIES; DEFIBRILLATOR/PACEMAKER Back to Search Results
Model Number MSERIES CCT
Medical Device Problem Codes Unable to Obtain Readings (1516); Unexpected Shutdown (4019)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional pro code: mkj.Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
 
Event or Problem Description
Complainant alleged that during functional testing, the device displayed a "poor pad contact" message and inappropriately shut down.Complainant did not indicate that there was any patient involvement in the reported malfunction.
 
Additional Manufacturer Narrative
The device was returned to zoll medical (b)(4); the reported malfunction for device shut down was duplicated and the ac power cord was replaced to remedy the problem.The device was recertified and returned to the customer.The reported malfunction for device displayed "poor pad contact" was evaluated and communication with the customer indicated there were no electrode pads attached at the time.This is normal operation of the device if there are no pads connected to the device.No trend is associated with reports of this type.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M SERIES
Common Device Name
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
MDR Report Key9177476
Report Number1220908-2019-03024
Device Sequence Number3447763
Product Code DPS
UDI-Device Identifier00847946014421
UDI-Public00847946014421
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K011865
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberMSERIES CCT
Device Catalogue NumberM SERIES
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 09/18/2019
Supplement Date Received by Manufacturer09/18/2019
Initial Report FDA Received Date10/10/2019
Supplement Report FDA Received Date11/11/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-