Model Number U1100 |
Device Problem
High Test Results (2457)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 09/09/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
A new test strip tray was sent to the reporter and upon installation of this new test strip tray on the urisys 1100 analyzer, no further issues with nitrite results were observed.This event occurred in (b)(6).
|
|
Event Description
|
The initial reporter stated they have been receiving false positive nitrite results for an unspecified number of patient samples tested using urisys 1100 analyzer serial number (b)(4).The issue started occurring after the urisys 1100 software was updated to version 5.71.When an affected sample is run on a test strip on the urisys 1100 analyzer, the nitrite result will be positive.When visually reading the same test strip, no color change is visible on the nitrite test pad.The issue occurs with different patient urine samples.The lot number and expiration date of the test strips used on the analyzer were requested, but not provided.
|
|
Manufacturer Narrative
|
The customer's urisys 1100 analyzer, tray, qc tray and retention materials showed no false positive results.The customer's urisys 1100 analyzer showed no defects and meets roche specifications.Led-intensity shows no significant deviation compared to reference device.The strip tray initially used by the customer may have been dirty or strips may have been positioned incorrectly.No product problem was identified.
|
|