ANIMAS LLC ONETOUCHPING GLUCOSEMGMTSYSTEM; INSULIN INFUSION PUMP Back to Search Results
Medical Device Problem Codes Electrical /Electronic Property Problem (1198); Structural Problem (2506)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluation: the device has been returned and evaluated by product analysis on 08-aug-2019 with the following findings: a review of the pump black box showed no power interruptions.There was no data in the black box from the time of the reported power issue due to continued use.A visual inspection of the pump revealed the battery compartment was cracked.The battery cap was not returned.A test battery cap was able to fit securely to maintain an electrical connection.The pump powered on with the test cap and was exercised for 12 hours with no power interruptions occurring.The pump¿s cover was removed and no damage was observed to the power printed circuit board.The complaint of intermittent power was not duplicated during investigation.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.Report late due to transition from vmsr program.
 
Event or Problem Description
On (b)(6) 2018 the reporter contacted animas, alleging a power (intermittent power) issue.There was no indication that the product caused or contributed to an adverse event.This complaint is being reported because the user may be unaware that the pump has lost power, leading to under delivery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ONETOUCHPING GLUCOSEMGMTSYSTEM
Common Device Name
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS LLC
965 chesterbrook blvd
wayne PA 19087
MDR Report Key9178623
Report Number2531779-2019-05384
Device Sequence Number11405120
Product Code LZG
UDI-Device Identifier10840406101034
UDI-Public10840406101034
Combination Product (Y/N)Y
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K080639
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/06/2019
Was the Report Sent to FDA? Yes
Device Age31 MO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/08/2019
Initial Report FDA Received Date10/11/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age52 YR
Patient Weight120
-
-