| Medical Device Problem Codes |
Electrical /Electronic Property Problem (1198); Structural Problem (2506)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluation: the device has been returned and evaluated by product analysis on 08-aug-2019 with the following findings: a review of the pump black box showed no power interruptions.There was no data in the black box from the time of the reported power issue due to continued use.A visual inspection of the pump revealed the battery compartment was cracked.The battery cap was not returned.A test battery cap was able to fit securely to maintain an electrical connection.The pump powered on with the test cap and was exercised for 12 hours with no power interruptions occurring.The pump¿s cover was removed and no damage was observed to the power printed circuit board.The complaint of intermittent power was not duplicated during investigation.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.Report late due to transition from vmsr program.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018 the reporter contacted animas, alleging a power (intermittent power) issue.There was no indication that the product caused or contributed to an adverse event.This complaint is being reported because the user may be unaware that the pump has lost power, leading to under delivery.
|
| |
|