Device evaluated by mfr.: returned product consisted of an emerge balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There were numerous kinks.There was a complete separation at 78.6cm distal of the strain relief.The balloon was tightly folded.The device was prepped with an inflation device filled with water and connected to an inflation port attached to the distal portion of the shaft separation to inflate the device.No issues were detected.The device was functionally tested with a.014 inch guidewire.No issues were detected.Inspection of the remainder of the device presented no other damage or irregularities.
|