| Model Number |
383401 |
| Medical Device Problem Code |
Improper Flow or Infusion (2954)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/02/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The incident sample has been requested but to date has not been received for evaluation.If the sample is received, or if additional information pertinent to the incident is obtained a follow-up report will be submitted.As part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.
|
| |
|
Event or Problem Description
|
|
The customer reported that the device is continuously feeding causing an over delivery outside of the accuracy limit.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary all device history records are reviewed for quality inspections and parameter compliance prior to releasing the product for shipment.The device was returned and evaluated at the regional service center.The device was tested for the reported condition and passed all testing.The reported issue could not be confirmed.A root cause could not be determined based on the device return and no corrective action is necessary at this time.Complaint trending information is reviewed on a routine basis and if a trend is observed, actions will be taken as necessary.
|
| |
|