BAYER MEDICAL CARE STELLANT CT SYRINGE KIT; ANGIOGRAPHIC SYRINGE KIT Back to Search Results
Model Number 60728737
Medical Device Problem Code Manufacturing, Packaging or Shipping Problem (2975)
Health Effect - Clinical Code Not Applicable (3189)
Date of Event 09/18/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Bayer product analysis received and examined the returned spike.Visual examination confirmed the presence of a foreign material under the blue dust cap.Our investigation determined that the foreign material was most likely introduced during the manufacturing process at the third-party supplier and went undetected.A 12-month review of complaint history determined the complaint rate to be 0.43 complaints per million (cpm).
 
Event or Problem Description
The customer reported the following: after opening the stellant ct disposable kit the customer observed a foreign material under the blue dust cap on the spike.No injury or adverse event was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STELLANT CT SYRINGE KIT
Common Device Name
ANGIOGRAPHIC SYRINGE KIT
Manufacturer (Section D)
BAYER MEDICAL CARE
1 bayer drive
indianola PA 15051
MDR Report Key9180011
Report Number2520313-2019-00044
Device Sequence Number2568644
Product Code DXT
UDI-Device Identifier00616258005188
UDI-Public(01)00616258005188
Combination Product (Y/N)N
Initial Reporter CountryNZ
PMA/510(K) Number
K082905
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 10/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/03/2023
Device Model Number60728737
Device Catalogue NumberSDS-CTP-SPK
Device Lot Number8560471
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/01/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/20/2019
Initial Report FDA Received Date10/11/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-