| Model Number |
M00573050 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/17/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a jagtome rx 44 was used in the duodenal papilla during an endoscopic retrograde cholangiopancreatography procedure performed on (b)(6), 2019.According to the complainant, during the procedure, the cutting wire broke prior to bowing the device.Reportedly, there was no part of the device detached inside the patient.The procedure was completed with a second jagtome rx 44, using the same generator and active cord.No complications were reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|
| |
|
Additional Manufacturer Narrative
|
|
(device codes): the problem code 1069 captures the reportable event of cutting wire broken.Visual examination of the return device revealed that the cutting wire was broken, bent and detached.There was no other issue noted.The complaint was consistent with the reported event of broken cutting wire.It is most likely that a peak of voltage could have caused the failures noted or if the device was not in contact with the tissue when it was energized.Therefore, the most probable cause of this complaint is adverse event related to procedure since it is the most likely that the adverse event occurred during the procedure and the device had no influence on event.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications at the time of release to distribution.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a jagtome rx 44 was used in the duodenal papilla during an endoscopic retrograde cholangiopancreatography procedure performed on (b)(6) 2019.According to the complainant, during the procedure, the cutting wire broke prior to bowing the device.Reportedly, there was no part of the device detached inside the patient.The procedure was completed with a second jagtome rx 44, using the same generator and active cord.No complications were reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|
| |
|