Model Number 24686
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/18/2019
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that shaft hole/perforated occurred.The target lesion was located in the right arm.A 12.0x40mm gladiator elite balloon catheter was advanced for dilatation.However, as the balloon was pulled negative, backflow of blood in the inflation syringe was noted due to a pinhole somewhere in the catheter.The procedure was completed with another of same device.No patient complications were reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: gladiator elite 8.0 x 40, 75 cm device was returned for analysis.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.Blood was observed in the balloon body indicating presence of a leak.The returned device was attached to an encore inflation unit and positive pressure was applied to inflate the balloon.The balloon could not be inflated due to the presence of solidified media that was present within the inflation lumen.The device was soaked in a water bath at a temperature of 37 degrees celsius to help soften the media before further inflation attempts were made.The returned device was again attached to a boston scientific encore inflation unit and positive pressure was applied to inflate the balloon and liquid was observed to be leaking from a balloon pinhole located at approximately 9 mm from the distal end of the proximal markerband.The rated burst pressure for this device is 20 atmospheres.A visual and microscopic examination observed no issues with the tip of the device which could potentially have contributed to the complaint incident.A visual and tactile examination found no kinks or damage along the shaft of the device.No other issues were identified during the product analysis.
 
Event or Problem Description
It was reported that shaft hole/perforated occurred.The target lesion was located in the right arm.A 12.0x40mm gladiator elite balloon catheter was advanced for dilatation.However, as the balloon was pulled negative, backflow of blood in the inflation syringe was noted due to a pinhole somewhere in the catheter.The procedure was completed with another of same device.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLADIATOR ELITE
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key9180180
Report Number2134265-2019-12019
Device Sequence Number11733651
Product Code LIT
UDI-Device Identifier08714729809876
UDI-Public08714729809876
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K132810
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number24686
Device Catalogue Number24686
Device Lot Number0024028510
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/23/2019
Initial Date Received by Manufacturer 09/19/2019
Supplement Date Received by Manufacturer11/08/2019
Initial Report FDA Received Date10/11/2019
Supplement Report FDA Received Date12/03/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age57 YR
-
-