| Model Number |
24686 |
| Medical Device Problem Code |
Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/18/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that shaft hole/perforated occurred.The target lesion was located in the right arm.A 12.0x40mm gladiator elite balloon catheter was advanced for dilatation.However, as the balloon was pulled negative, backflow of blood in the inflation syringe was noted due to a pinhole somewhere in the catheter.The procedure was completed with another of same device.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: gladiator elite 8.0 x 40, 75 cm device was returned for analysis.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.Blood was observed in the balloon body indicating presence of a leak.The returned device was attached to an encore inflation unit and positive pressure was applied to inflate the balloon.The balloon could not be inflated due to the presence of solidified media that was present within the inflation lumen.The device was soaked in a water bath at a temperature of 37 degrees celsius to help soften the media before further inflation attempts were made.The returned device was again attached to a boston scientific encore inflation unit and positive pressure was applied to inflate the balloon and liquid was observed to be leaking from a balloon pinhole located at approximately 9 mm from the distal end of the proximal markerband.The rated burst pressure for this device is 20 atmospheres.A visual and microscopic examination observed no issues with the tip of the device which could potentially have contributed to the complaint incident.A visual and tactile examination found no kinks or damage along the shaft of the device.No other issues were identified during the product analysis.
|
| |
|
Event or Problem Description
|
|
It was reported that shaft hole/perforated occurred.The target lesion was located in the right arm.A 12.0x40mm gladiator elite balloon catheter was advanced for dilatation.However, as the balloon was pulled negative, backflow of blood in the inflation syringe was noted due to a pinhole somewhere in the catheter.The procedure was completed with another of same device.No patient complications were reported.
|
| |
|
Search Alerts/Recalls
|