| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
09/13/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being filed to provide additional information.Investigation: a dhr search was not performed for this complaint record as the reported adverse event did not relate to activities associated with the manufacturing of this lot.All lots must meet acceptance criteria for release.Correction: terumo bct customer support contacted the distributor on (b)(6) 2019 to alert them of this error.Root cause: based on the clinical findings, the use of the expired set was caused by an operator error.
|
| |
|
Event or Problem Description
|
|
The customer did not provide any patient information or outcome.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided.
|
| |
|
Event or Problem Description
|
|
Upon review of the information provided by the distributor, it was discovered that an expired gdp-10 graft delivery system disposable was used on a patient.The gdp-10 disposable was labeled with an expiration date of 09-01-2019 and the procedure was performed on (b)(6) 2019.The product was collected and processed then injected back to the patient.Patient information and outcome are not available at this time.The gdp-10 graft delivery system set is not available for return because it was discarded by the customer.
|
| |
|