MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE CONV FB PS TB TRL SZ6; ATTUNE INSTRUMENTS : TIBIAL TRIALS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE CONV FB PS TB TRL SZ6; ATTUNE INSTRUMENTS : TIBIAL TRIALS Back to Search Results
Model Number 2545-00-976
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Not Applicable (3189)
Date of Event 09/20/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that when removing the size 6 attune fb ps trial, the posterior corner broke off into 2 pieces.All pieces were retrieved.1 minute surgical delay.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Added: common device name, device identification, concomitant medical products, pma/510k and device manufacture date.Corrected: device evaluated by mfr.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: product complaint (b)(4).Investigation summary : examination of the returned device confirmed the reported breakage.Root cause and corrective action are documented in the capa system.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Corrected d4 (expiration date): no expiration date to be reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE CONV FB PS TB TRL SZ6
Common Device Name
ATTUNE INSTRUMENTS : TIBIAL TRIALS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key9180899
Report Number1818910-2019-109409
Device Sequence Number11532517
Product Code JWH
UDI-Device Identifier10603295135531
UDI-Public10603295135531
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K140881
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 09/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2545-00-976
Device Catalogue Number254500976
Device Lot NumberBFA0MSV
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/31/2019
Initial Date Received by Manufacturer 09/20/2019
Supplement Date Received by Manufacturer11/14/2019
11/14/2019
01/24/2020
Initial Report FDA Received Date10/11/2019
Supplement Report FDA Received Date11/11/2019
11/14/2019
01/30/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age50 YR
-
-