• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US PFCSIG FLUT FEMROD 5DG18X125MM; SIGMA REVISION IMPLANT : KNEE STEMS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US PFCSIG FLUT FEMROD 5DG18X125MM; SIGMA REVISION IMPLANT : KNEE STEMS Back to Search Results
Catalog Number 960703
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Pain (1994); No Code Available (3191)
Date of Event 09/20/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Reporting a revision of sigma tc3 fixed bearing knee implants at royal lancaster infirmary.All components were removed without difficulty.It should be noted that upon removal of the femoral component, it was observed that the femoral component and the femoral adaptor were slightly loose and can be rotated ¿ questioning whether insufficient torque was used upon initial implant preparation by surgical team.The disassociation occurred between the bolt and the stem as this is a stem-bolt configuration.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary
=
> the devices were reviewed by bioengineering and a report was received stating it is unlikely that a potential product issue was present root cause undetermined.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthese has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthese or its employees that the report constitutes an admission that the device, depuy synthese, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PFCSIG FLUT FEMROD 5DG18X125MM
Common Device Name
SIGMA REVISION IMPLANT : KNEE STEMS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9180949
Report Number1818910-2019-109410
Device Sequence Number2774779
Product Code JWH
UDI-Device Identifier10603295233640
UDI-Public10603295233640
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K952830
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2019
Reporter Type Manufacturer
Report Source foreign,health professional,o
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number960703
Device Lot NumberFH6LP4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/18/2019
Initial Date Received by Manufacturer 09/20/2019
Supplement Date Received by Manufacturer10/04/2019
11/19/2019
Initial Report FDA Received Date10/11/2019
Supplement Report FDA Received Date10/21/2019
11/29/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
P.F.C.* FLUT TIB ROD 75X16MM; PFC SIGMA MOD TIB TRAY CEM SZ4; PFC*SIGMA DI AUG12MMCO,SZ5RGHT; PFC*SIGMA DIS AUG 8MM,SZ5,RGHT; SIGMA PS CEM FEM SZ5 R; SIGMA STAB GVF INS 4 15MM
Outcome Attributed to Adverse Event Required Intervention;
-
-