• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. FLUENCY PLUS ENDOVASCULAR STENT GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. FLUENCY PLUS ENDOVASCULAR STENT GRAFT Back to Search Results
Model Number FEM14040
Medical Device Problem Codes Positioning Failure (1158); Material Perforation (2205); Misfire (2532)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/28/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturing review: review of manufacturing records was not required, as no additional complaint has been reported for this lot.Based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.However, the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met all specifications prior to shipment.Investigation summary: a stent graft delivery system was returned for evaluation.On the basis of the returned stent graft delivery system, the alleged deployment failure could be confirmed.The stent graft was found partially released and the outer sheath was found elongated, which indicated that increased friction affected the delivery system during attempt of stent graft deployment.Furthermore, the outer sheath was found perforated by stent graft strut, which made a stent graft deployment impossible.As a result of the investigation performed the complaint is confirmed.A definite root cause for the reported event could not be determined.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.
 
Event or Problem Description
It was reported that during a stent graft placement , the stent allegedly failed to deploy.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLUENCY PLUS ENDOVASCULAR STENT GRAFT
Common Device Name
ENDOVASCULAR STENT GRAFT
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key9181671
Report Number2020394-2019-03864
Device Sequence Number2774217
Product Code PFV
UDI-Device Identifier04049519008684
UDI-Public(01)04049519008684
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
P130029
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 10/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/09/2021
Device Model NumberFEM14040
Device Catalogue NumberFEM14040
Device Lot NumberANCQ1363
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/24/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/24/2019
Initial Report FDA Received Date10/11/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age58 YR
Patient Weight106
-
-