ZOLL MEDICAL CORPORATION X SERIES; DEFIBRILLATOR/PACEMAKER Back to Search Results
Model Number X SERIES
Medical Device Problem Code Unable to Obtain Readings (1516)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/20/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
 
Event or Problem Description
Complainant alleged that while attempting to monitor an 89 year-old-male patient during cardiac arrest, the device was unable to obtain an ecg signal via electrode pads.Complainant indicated that the clinician obtained another device to continue treating the patient.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.
 
Additional Manufacturer Narrative
Zoll medical corporation evaluated the device and the customer's report was not replicated or confirmed.Review of the device logs did show the ecg signal was visible when the pads were attached and the device was in pads view.The logs also show the view switches to lead i and the signal is lost.The patient impedance waveform has a solid reading at this time, confirming the device recognized the impedance through pads.The user stated that they attempted to resolve the issue by disconnecting and reconnecting cables and leads, but there is no evidence of leads being attached to the patient; only pads.It is noteworthy to mention the electrode pads used at the time of the reported event were not returned for evaluation.Analysis of reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X SERIES
Common Device Name
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
MDR Report Key9181857
Report Number1220908-2019-03033
Device Sequence Number2591550
Product Code DRT
UDI-Device Identifier00847946018276
UDI-Public00847946018276
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K112432/P160
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Other
Device Model NumberX SERIES
Device Catalogue NumberX SERIES
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/30/2019
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 09/20/2019
Supplement Date Received by Manufacturer09/20/2019
Initial Report FDA Received Date10/11/2019
Supplement Report FDA Received Date12/09/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age89 YR
-
-