| Catalog Number |
RF621 SMN 10464525 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/17/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Mdrs 2517506-2019-00389 and 2517506-2019-00390 were filed for the same event.The customer contacted the siemens customer care center (ccc) and reported discordant falsely elevated cardiac troponin i (tni) patient results were obtained on a dimension exl 200 instrument.Siemens healthcare diagnostics headquarters support center (hsc) completed the investigation of the event.Hsc has reviewed the information provided and the instrument data.No product non-conformance was identified.The patient sample was requested be returned to the siemens technical support laboratory for evaluation.The returned sample produced a troponin value of 0.113 ng/ml which did not change when treated with the heterophile blocking pre-treatment (hbt).The data shows that the patient produces a repeatedly elevated result with the dimension exl tni assay that is not consistent with alternate non-siemens methodologies and clinical presentation of the patient.The cause of the event in is unknown.The system is operational.The device is performing within specifications.No further evaluation is required.
|
| |
|
Event or Problem Description
|
|
A discordant, elevated cardiac troponin i (tni) result was obtained on a patient serum/plasma sample on a dimension exl 200 instrument.The result was reported to the physician(s) who questioned the result.The same sample was reprocessed on the same instrument and higher results were obtained; the higher results were not reported to the physician(s).A new sample was drawn from the same patient and processed on an alternate non-siemens instrument in an alternate facility; a lower result was obtained which was considered correct.There are no known reports of patient intervention or adverse health consequences due to the discordant tni results.
|
| |
|
Search Alerts/Recalls
|