ABBOTT VASCULAR XIENCE XPEDITION; DRUG ELUTING CORONARY STENT DELIVERY SYSTEM Back to Search Results
Catalog Number 1074350-28
Medical Device Problem Codes Material Separation (1562); Tear, Rip or Hole in Device Packaging (2385); Deformation Due to Compressive Stress (2889); Packaging Problem (3007)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 09/20/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Exemption number e2019001.The device is expected to be returned for evaluation.It has not yet been received.A follow up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that during unpacking of a 3.50x28mm xience xpediition stent delivery system, the outer packaging was noted with a slight dent.As the device was being unpacked further damage was noted (dispenser hoop coil dented, delivery system (shaft) was kinked and the stent implant [broken] in two pieces).The device was never used and there was no patient involvement.Another xience xpedition was used to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
The device was returned for analysis.The reported material separation, deformation due to compressive stress (shaft kink), and packaging problem (kinked coil) were confirmed.The reported tear, rip, or hole in device packaging (chipboard box damage) could not be confirmed as the box was inadvertently discarded during inventory receiving.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.The investigation determined the reported difficulties appear to be related to operational context.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event or Problem Description
It was reported that during unpacking of a 3.50x28mm xience xpediition stent delivery system, the outer packaging was noted with a slight dent.As the device was being unpacked further damage was noted (dispenser hoop coil dented, delivery system (shaft) was kinked and the stent implant [broken] in two pieces).The device was never used and there was no patient involvement.Another xience xpedition was used to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.(b)(6) 2019: per reassessment of the event details, it was confirmed that the hypotube separated in two pieces, not the stent implant.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE XPEDITION
Common Device Name
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key9182666
Report Number2024168-2019-12569
Device Sequence Number2615562
Product Code NIQ
UDI-Device Identifier08717648162565
UDI-Public08717648162565
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/18/2022
Device Catalogue Number1074350-28
Device Lot Number9031241
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/14/2019
Initial Date Received by Manufacturer 09/20/2019
Supplement Date Received by Manufacturer11/06/2019
Initial Report FDA Received Date10/11/2019
Supplement Report FDA Received Date11/20/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No