Catalog Number 1074350-28 |
Device Problems
Material Separation (1562); Tear, Rip or Hole in Device Packaging (2385); Deformation Due to Compressive Stress (2889); Packaging Problem (3007)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 09/20/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Exemption number e2019001.The device is expected to be returned for evaluation.It has not yet been received.A follow up report will be submitted with all additional relevant information.
|
|
Event Description
|
It was reported that during unpacking of a 3.50x28mm xience xpediition stent delivery system, the outer packaging was noted with a slight dent.As the device was being unpacked further damage was noted (dispenser hoop coil dented, delivery system (shaft) was kinked and the stent implant [broken] in two pieces).The device was never used and there was no patient involvement.Another xience xpedition was used to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
|
|
Manufacturer Narrative
|
The device was returned for analysis.The reported material separation, deformation due to compressive stress (shaft kink), and packaging problem (kinked coil) were confirmed.The reported tear, rip, or hole in device packaging (chipboard box damage) could not be confirmed as the box was inadvertently discarded during inventory receiving.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.The investigation determined the reported difficulties appear to be related to operational context.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|
|
Event Description
|
It was reported that during unpacking of a 3.50x28mm xience xpediition stent delivery system, the outer packaging was noted with a slight dent.As the device was being unpacked further damage was noted (dispenser hoop coil dented, delivery system (shaft) was kinked and the stent implant [broken] in two pieces).The device was never used and there was no patient involvement.Another xience xpedition was used to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.(b)(6) 2019: per reassessment of the event details, it was confirmed that the hypotube separated in two pieces, not the stent implant.No additional information was provided.
|
|