• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN LP LLC NORTH HAVEN VERISET; AGENT, ABSORBABLE HEMOSTATIC, NON-COLLAGEN BASED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LP LLC NORTH HAVEN VERISET; AGENT, ABSORBABLE HEMOSTATIC, NON-COLLAGEN BASED Back to Search Results
Model Number HP0510E
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Post Operative Wound Infection (2446); No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This product is not sold in us and is 510(k) exempt.(b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, post-operatively a total thyroidectomy procedure, the patient had a wound infection after using the patch.The patch was opened from the foil pouch and placed on the sterile field an hour prior to being used.Patient's hospitalization was extended and was provided with iv antibiotics.
 
Additional Manufacturer Narrative
Post market vigilance (pmv) led an evaluation two photographs of a device on the throat of a patient with evident swelling and redness at the site.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.A review of the device history record indicates the product was released meeting all quality release specifications at the time of manufacture.Without the physical product a root cause could not be determined.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERISET
Common Device Name
AGENT, ABSORBABLE HEMOSTATIC, NON-COLLAGEN BASED
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
MDR Report Key9182963
Report Number1219930-2019-05736
Device Sequence Number16834978
Product Code LMG
UDI-Device Identifier20884521082554
UDI-Public20884521082554
Combination Product (Y/N)N
Initial Reporter CountryAS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/18/2020
Device Model NumberHP0510E
Device Catalogue NumberHP0510E
Device Lot NumberN9E1016Y
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/22/2019
Supplement Date Received by Manufacturer01/10/2020
Initial Report FDA Received Date10/11/2019
Supplement Report FDA Received Date01/13/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-