| Model Number |
654401 |
| Medical Device Problem Code |
Particulates (1451)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
09/01/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is expected to return.It has not been received.
|
| |
|
Event or Problem Description
|
|
The event occurred on an unknown date and involved a latex free cystoscopy irrigation set that was reported to have a hair inside of the tubing noted while still in the packaging.There was no patient involved.
|
| |
|
Additional Manufacturer Narrative
|
|
No list #065440403 product samples or videos were returned for investigation.An image was returned which shows a hair on the set, it is not clear from the image whether the hair was inside of the package, on the tubing, or within the tubing (in the fluid path).The reported complaint can be confirmed, the probable cause is due to a gowning error at the manufacturing site.The lot history was reviewed and there were no nonconformance's found that would have contributed to the reported complaint.
|
| |
|