| Catalog Number |
6260-9-040 |
| Medical Device Problem Codes |
Malposition of Device (2616); Noise, Audible (3273)
|
| Health Effect - Clinical Codes |
Arthritis (1723); Pain (1994)
|
| Date of Event |
09/16/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Lot specific voluntary recall was initiated for the lfit v40 cocr heads within scope of a capa.The investigation revealed that only specific catalog numbers and specific lots are impacted by the regulatory action and that the affected lots were manufactured on or before march 4, 2011.The root cause analysis identified a process related anomaly as to the affected sizes and lots.The affected product has all been implanted and/or expired.Device not returned.
|
| |
|
Event or Problem Description
|
|
Patient returned to dr's office to ask the pain in his left hip due to arthritis.During this appointment the patient stated his right hip has started clicking but was not painful.While examining the radiograph dr noticed the trunnions was not in the center of the implant head.It looked eccentric.A revision surgery was then scheduled.
|
| |
|
Search Alerts/Recalls
|