The conmed representative reported on behalf of the facility that the 2001m-c, pad pro defib pads were used in the catherization lab during a procedure on (b)(6) 2019.They had to use multiple pads from a couple of different lots and experienced no ekg readings.The pads were replaced with competitor pads and tracing was performed as normal.The conmeds representative was notified at the end of the day and immediately responded with no return communication.He stopped onsite the next morning to gather samples, and intel of situation and what is going to happen.The current status of the patient is stable.This report is being raised based on device malfunction with potential for injury upon reoccurrence.
|
|
Received one product sample in original, unopened packaging and 2 empty pouches.Evaluation of the device from the unopened packaging found no issues with electrical continuity, peel functionality or visual attributes.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution were found to have met all specifications prior to shipment.This is the only complaint for this lot number and failure mode within the past two years.(b)(4).Per the instructions for use, the user is advised the following; - make sure the skin is dry i.E.Free of water, sweat, alcohol, etc.- do not apply any substance to the skin surface that will leave a residue, i.E.Lotions, oils, etc.- do not fold, trim, crush, or store under heavy objects.- misuse of multifunction electrodes, use of compromised or altered product, and/or failure to follow product instructions may result in patient burns, inadequate delivery of therapy, and/or loss of ecg trace quality.This issue will continue to be monitored through the complaint system to assure patient safety.
|