Catalog Number 031.034S
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Codes Abscess (1690); Hemorrhage/Blood Loss/Bleeding (1888); Inflammation (1932); Pain (1994); Swelling (2091)
Date of Event 08/30/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The batch number could be verified.Our manufacturing q-system assures, that production and process controls are in place to ensure that batches confirm to the applicable specifications before they are distributed.The non-integration of an endosseous dental implant during the healing phase is a known inherent risk of the treatment with dental implants.Most implant failures occur before occlusal loading (analainen et al.2013, koldsland et al.2009).Based on clinical studies about 1-3% of the implants fail within the first year after implantation (ganeles et al.2008).Implants may have to be removed in case one or more of the implant success criteria are not met.Implant success criteria according to buser et al.(1991) are: absence of persistent subjective complaints such as pain, foreign body sensation and /or disesthesia.Absence of a recurrent peri-implant infection with suppuration.Absence of implant mobility.Absence of a continuous radiolucency around the implant.The manufacturer's trend analysis confirms that the reported early failure rate associated with its dental implants is below the expected failure rate for this treatment as published in the scientific literature.
 
Event or Problem Description
The clinician reported that two weeks after the dental implant was placed, in ada site #8 of the patient's mouth, non-osseointegration was observed.The patient presented type iv bone.The device was returned to the manufacturer for investigation.Patient experienced pain, bleeding, swelling, inflammation and abscess at time of event, no further complications were observed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CIM Ø4.1MM RD, ZLA 14MM, AH5.5MM, ZRO2
Common Device Name
ENDOSSEOUS DENTAL IMPLANT
Manufacturer (Section D)
INSTITUT STRAUMANN AG
peter merian-weg 12
basel, CH-40 02
SZ  CH-4002
MDR Report Key9189043
Report Number1222315-2019-09790
Device Sequence Number11733684
Product Code DZE
UDI-Device Identifier07630031714015
UDI-Public07630031714015
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K171769
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 10/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/05/2023
Device Catalogue Number031.034S
Device Lot NumberRH924
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/20/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/20/2019
Initial Report FDA Received Date10/14/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age48 YR
-
-