• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC SINGAPORE OPERATIONS CAPSURE SENSE LEAD MRI SURESCAN; ELECTRODE, PACEMAKER, PERMANENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC SINGAPORE OPERATIONS CAPSURE SENSE LEAD MRI SURESCAN; ELECTRODE, PACEMAKER, PERMANENT Back to Search Results
Model Number 407458
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 09/17/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient's pacemaker incision did not heal properly resulting in the implantable pulse generator (ipg) being exposed.It was further reported that the patient developed a pocket infection.The ipg system was explanted and replaced six weeks after implant.No further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPSURE SENSE LEAD MRI SURESCAN
Common Device Name
ELECTRODE, PACEMAKER, PERMANENT
Manufacturer (Section D)
MEDTRONIC SINGAPORE OPERATIONS
49 changi south avenue 2
nasaco tech centre
singapore 48605 6
SG  486056
MDR Report Key9189871
Report Number3008973940-2019-02812
Device Sequence Number1840500
Product Code DTB
UDI-Device Identifier00643169448957
UDI-Public00643169448957
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
P830061
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date10/03/2019
Device Model Number407458
Device Catalogue Number407458
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/17/2019
Initial Report FDA Received Date10/15/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age72 YR
-
-