• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 PUREWICK DRYDOC STATION; DRY DOC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 PUREWICK DRYDOC STATION; DRY DOC Back to Search Results
Catalog Number DD15
Medical Device Problem Code Increase in Suction (1604)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); Tissue Breakdown (2681)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Event or Problem Description
It was reported that there was a skin breakdown while using the dry doc.
 
Additional Manufacturer Narrative
The reported event was inconclusive as no sample was returned for evaluation.The potential root cause could be a defective pump.He lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "replace the purewick female external catheter at least every 8-12 hours or if soiled with feces or blood.Always assess skin for compromise and perform perineal care prior to placement of a new purewick female external catheter".Correction: d4 h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The device was not returned.
 
Event or Problem Description
It was reported that there was a skin breakdown while using the dry doc.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PUREWICK DRYDOC STATION
Common Device Name
DRY DOC
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key9190983
Report Number1018233-2019-06455
Device Sequence Number2107323
Product Code NZU
UDI-Device Identifier00801741151781
UDI-Public(01)00801741151781
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue NumberDD15
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/25/2019
Supplement Date Received by Manufacturer10/23/2019
Initial Report FDA Received Date10/15/2019
Supplement Report FDA Received Date10/28/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-