Model Number 204990
Medical Device Problem Codes Fluid/Blood Leak (1250); Contamination of Device Ingredient or Reagent (2901)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/02/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As part of the normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.
 
Event or Problem Description
Piece is leaking oil.Case type: pka (mics).Patient under anesthesia.
 
Event or Problem Description
Piece is leaking oil.Case type: pka (mics).Patient under anesthesia.
 
Additional Manufacturer Narrative
Reported event: it was reported that piece is leaking oil case type: pka (mics) patient under anesthesia.Product evaluation and results: the product was not evaluated as the product was unavailable for inspection capa 2127499 has been raised for the same.Product history review: review of the device history records indicate 50 devices were manufactured and 49 accepted (including serial 4001390) into final stock on 06/17/2019 with no reported discrepancies.Review of qt 19-06-0080 revealed that the non-conformance is unrelated to the failure alleged in the event.Complaint history review: a review of complaints in catsweb and trackwise related to p/n 204990, lot 40m115490 shows 0 additional complaints related to the failure in this investigation.Conclusions: the failure could not be determined as the product was not available for inspection.No additional investigation or specific actions are required.If additional information is received then the complaint will be reopened.Corrective action/preventive action: a review of stryker¿s nc/capa database indicated there have been no nc and capa associated with the product and failure mode reported in this event.H3 other text : device not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH SPEED ROTARY ATTACHMENT
Common Device Name
STEREOTAXIC DEVICE, ROBOTICS
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
MDR Report Key9191239
Report Number3005985723-2019-00734
Device Sequence Number2599447
Product Code OLO
UDI-Device Identifier00848486021849
UDI-Public00848486021849
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K142530
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number204990
Device Catalogue Number204990
Device Lot Number4001390
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 10/02/2019
Supplement Date Received by Manufacturer02/22/2020
Initial Report FDA Received Date10/15/2019
Supplement Report FDA Received Date03/11/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
-
-