|
Discordant, falsely high cardiac troponin (tni) patient results were obtained on a dimension exl 200 instrument.The initial result for sid (b)(6) was reported to the physician(s) and questioned.A new sample was drawn under sid (b)(6), on the same patient and repeated on the same instrument, resulting in a lower value.Additionally, both samples were repeated on the same instrument, resulting in a lower value.The initial results for sid (b)(6) and (b)(6) were reported to the physician(s) and questioned.The same samples were repeated on the same instrument, resulting in lower values.The repeat results were reported, as the correct results, to the physician(s).It was reported that a patient (sid (b)(6)) received unnecessary medication as a result.There are no reports of patient intervention or adverse health consequences due to the discordant results.
|