| Model Number |
620RG29 |
| Medical Device Problem Code |
Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Codes |
Host-Tissue Reaction (1297); Mitral Valve Stenosis (1965)
|
| Date of Event |
12/21/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
No product was returned.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that 11 years and 1 month post implant of this 29mm mitral annuloplasty ring, it was explanted and replaced with a non-medtronic mechanical valve.The reason for replacement was reported as severe mitral stenosis.It was also noted there was pannus overgrowth on the patient's native mitral valve.No additional adverse patient effects were reported.
|
| |
|