| Model Number |
STRIP, RLN TMX 50CTMG/DL |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/17/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Meter and test strips were not returned for evaluation.Most likely underlying root cause: mlc-12: product expired.Note: manufacturer contacted customer in a follow-up call to ensure the replacement products resolved the initial concern - unable to establish contact with customer at this time.Product notification letter sent to contact customer care.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for low blood glucose test results.The customer is concerned with tests results obtained of 47 mg/dl and 41 mg/dl.The expected fasting blood glucose test result range is 70 mg/dl - 138 mg/dl.The customer did not report symptoms.Medical attention is not reported as a result of the actual blood glucose results.The product is stored according to specification in the bedroom.During the call a back to back blood test was not performed by the customer.The test strip lot manufacturer¿s expiration date is 07/13/2019 and open vial date is undisclosed.The meter memory was reviewed for previous test result history.(b)(6).
|
| |
|