| Catalog Number |
109.327 |
| Medical Device Problem Code |
Loss of Osseointegration (2408)
|
| Health Effect - Clinical Codes |
Unspecified Infection (1930); Pain (1994); Tissue Damage (2104)
|
| Date of Event |
08/21/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Considering the surgical methodology, it was seen that the dentist has used more drills than recommended to perform the implant installation.The investigation of the complaint was carried out considering the analysis of the information given by the dentist in the guarantee form, visual conference of the product returned by the dentist and the evaluation of relevant production records.
|
| |
|
Event or Problem Description
|
|
(b)(4) - the dentist reported that 5 years and 5 months after the dental implant was installed in intraoral region 18#, its loss of osseointegration was observed.Moreover, 32n.Cm of primary stability was achieved and the patient presented bone type ii.
|
| |
|
Search Alerts/Recalls
|