| Model Number |
MT20649 |
| Medical Device Problem Code |
Defective Alarm (1014)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/01/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
The product was evaluated.An exterior visual inspection was performed and passed.Charged and boot were performed and passed.Receiver functional testing was performed and passed."try it" testing was performed and passed.Drop test for intermittent audio was performed and passed.Case opened for interior inspection was performed and passed.Audio speaker resistance was measured and passed.The receiver log was downloaded and reviewed finding no errors related to the complaint.The allegation was not confirmed.The probable cause could not be determined.No injury or medical intervention was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that intermittent audio output occurred.No product or data was provided for evaluation.Confirmation of the allegation and a probable cause could not be determined.No injury or medical intervention was reported.
|
| |
|