RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Mechanical Problem (1384)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2019.Date of report: 10/15/2019.The manufacturer¿s international service technician confirmed the reported pressure sensor issue.The manufacturer¿s international service technician replaced the data acquisition(da) printed circuit board assembly(pcba) to address the reported problem.The unit was checked overall, run in tested, cleaned and functionally tested and no abnormality was confirmed.
|
|
Event Description
|
Proximal pressure sensor auto-zero failed.There was no patient involvement.
|
|
Manufacturer Narrative
|
Mfr received date: 28 nov 2019.The part was returned to the manufacturer for failure investigation (fi).No fault found.Customer complaint could not be replicated.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|
|