ABBOTT VASCULAR NC TREK RX; CORONARY DILATATION CATHETER Back to Search Results
Catalog Number 1012445-12
Medical Device Problem Codes Break (1069); Material Separation (1562); Noise, Audible (3273)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/25/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Exemption number e2019001.The device was received.Investigation is not yet complete.A follow up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that the procedure was to treat a moderately calcified unspecified coronary artery.A 2.0 x 12 mm nc trek balloon catheter was used; however, it became "damaged/broken" during procedure.There were no adverse patient effects.No additional information was provided.
 
Additional Manufacturer Narrative
Visual inspection was performed on the returned device.The reported separation was confirmed.The reported popping noise could not be replicated in a testing environment as it was based on operational circumstances.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents/complaints from this lot.The investigation determined the reported separation appears to be related to operational context; however, a conclusive cause for the reported popping noise could not be determined.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported that the procedure was to treat a moderately calcified unspecified coronary artery.A 2.0 x 12 mm nc trek balloon catheter was used; however, it became "damaged/broken" during procedure.There were no adverse patient effects.Subsequent to the initial report, additional information was added.The device made a popping sound during a kissing balloon technique and the shaft of the nc trek balloon catheter separated in the hemostatic valve.The components were simply withdrawn.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK RX
Common Device Name
CORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key9194704
Report Number2024168-2019-12638
Device Sequence Number3935126
Product Code LOX
UDI-Device Identifier08717648151644
UDI-Public08717648151644
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K103153
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2019
Device Catalogue Number1012445-12
Device Lot Number70118G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/08/2019
Initial Date Received by Manufacturer 09/25/2019
Supplement Date Received by Manufacturer12/04/2019
Initial Report FDA Received Date10/15/2019
Supplement Report FDA Received Date12/04/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age71 YR
-
-