U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR GRAFTMASTER; CORONARY STENT DELIVERY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR GRAFTMASTER; CORONARY STENT DELIVERY SYSTEM Back to Search Results
Catalog Number 1012583-16
Medical Device Problem Code Device Damaged by Another Device (2915)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/28/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Exemption number e2019001.The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history identified no other incidents.The investigation determined that the reported device damaged by another device appears to be related to circumstances of the procedure.Based on the information reviewed, there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.The wiggle guide wire referenced was filed under a separate medwatch report number.
 
Event or Problem Description
Initial information received via user facility medwatch regarding a wiggle guide wire states: when attempted to remove wire, the tip broke off and became entrapped behind covered stent.Patient remained hemodynamically stable throughout event and recovery.Additional information received indicates that the procedure was to treat a free perforation in the mid right coronary artery (rca).The wiggle guide wire was trapped when a graftmaster covered stent was implanted.The perforation was sealed.When attempting to remove the wire there was significant resistance and the tip detached.The tip was embedded to the vessel wall by the use of an additional stent.The patient was discharged the following day.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GRAFTMASTER
Common Device Name
CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key9194893
Report Number2024168-2019-12643
Device Sequence Number2486487
Product Code MAF
UDI-Device Identifier08717648176470
UDI-Public08717648176470
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
H000001
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2020
Device Catalogue Number1012583-16
Device Lot Number8073141
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/24/2019
Initial Report FDA Received Date10/15/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age85 YR
Patient Weight77
-
-