| Catalog Number |
1012583-16 |
| Medical Device Problem Code |
Device Damaged by Another Device (2915)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/28/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Exemption number e2019001.The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history identified no other incidents.The investigation determined that the reported device damaged by another device appears to be related to circumstances of the procedure.Based on the information reviewed, there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.The wiggle guide wire referenced was filed under a separate medwatch report number.
|
| |
|
Event or Problem Description
|
|
Initial information received via user facility medwatch regarding a wiggle guide wire states: when attempted to remove wire, the tip broke off and became entrapped behind covered stent.Patient remained hemodynamically stable throughout event and recovery.Additional information received indicates that the procedure was to treat a free perforation in the mid right coronary artery (rca).The wiggle guide wire was trapped when a graftmaster covered stent was implanted.The perforation was sealed.When attempting to remove the wire there was significant resistance and the tip detached.The tip was embedded to the vessel wall by the use of an additional stent.The patient was discharged the following day.No additional information was provided.
|
| |
|