Model Number AT75186
Medical Device Problem Codes Difficult to Remove (1528); Retraction Problem (1536); Unraveled Material (1664)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Event or Problem Description
This report summarizes one malfunction.A review of the reported information indicates that model at75186, pta balloon dilation catheter, allegedly became caught in the stent, was difficult to remove from the patient, and experienced unraveled material.This information was received from a single source.This malfunction involved a patient with no consequences.Patient age, weight, and gender were not provided.
 
Event or Problem Description
This report summarizes one malfunction.A review of the reported information indicates that model at75186, pta balloon dilation catheter, allegedly became caught in the stent, was difficult to remove from the patient, and experienced unraveled material.This information was received from a single source.This malfunction involved a patient with no consequences.Patient age, weight, and gender were not provided.
 
Additional Manufacturer Narrative
H10: the lot number for the malfunction was provided and a lot history review was performed.The investigation is confirmed for the unraveled fibers and inconclusive for the retraction issue.The definitive root cause could not be determined.The device is labeled for single use.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATLAS PTA DILATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key9195157
Report Number2020394-2019-03980
Device Sequence Number3671736
Product Code LIT
UDI-Device Identifier00801741062506
UDI-Public(01)00801741062506
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model NumberAT75186
Device Catalogue NumberAT75186
Device Lot NumberGFCY2820
Initial Date Received by Manufacturer 09/30/2019
Supplement Date Received by Manufacturer12/31/2019
Initial Report FDA Received Date10/15/2019
Supplement Report FDA Received Date01/07/2020
Patient Sequence Number1
-
-