| Catalog Number |
03.50.030 |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
Pain (1994)
|
| Date of Event |
09/11/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Batch review performed on (b)(6) 2019; lot 155082: (b)(4) items manufactured and released on (b)(6) 2015.Expiration date: 2020-09-05.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold with another similar reported event.Two equal device were reported in this complaint.
|
| |
|
Event or Problem Description
|
|
The patient complained about pain during a control visit.Two pedicle screws were found to be broken at s1 level.Revision surgery was performed almost 2 years and a half after primary.The screws' threaded parts were left in the patient.The surgeon implanted a competitor system, the new screws in s1 were implanted with a parallel trajectory.The surgery was completed successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual inspection performed by medacta spine r&d project manager: the two pedicle screw ref.03.50.030, lot 155082, are broken approximately 10mm below the pedicle screw head.All the other explanted implant are without any damaging.The root cause of the event is unknown, probably is due to a huge load on the lower pedicle screw implanted in s1.Despite of the breakage as reported in the tf: vi-ps-01 the implant, in worst case configuration, can accomplish the requirements of the standard and it's comparable or better refer to the implants already on the market.
|
| |
|
Event or Problem Description
|
|
Device returned and analysed by medacta r&d.
|
| |
|