| Model Number |
MMT-722LNAL |
| Medical Device Problem Codes |
No Display/Image (1183); Device Difficult to Program or Calibrate (1496)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/11/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump alarm with button error as well as the insulin pump was not working.The customer blood glucose level was unknown.Customer was assisted with troubleshooting.Customer stated that the time clock was advance for one minute.Customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The device will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with no (act, up and escape) button response due to corroded keypad traces.No button error alarm and no blank display anomaly during testing noted.Unable to perform all functional testing including the displacement, operating currents, a21 error test, rewind, basic occlusion, occlusion, prime a33 and excessive no delivery test due to buttons not responding.
|
| |
|