MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH MRS FEM STEM W/O BODY 13X127MM; PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH MRS FEM STEM W/O BODY 13X127MM; PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Catalog Number 64853113
Medical Device Problem Code Osseointegration Problem (3003)
Health Effect - Clinical Codes Pain (1994); Injury (2348); Inadequate Osseointegration (2646)
Date of Event 09/19/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
 
Event or Problem Description
It was reported the patient's left knee was revised.As reported by rep: "patient presented with thigh pain, it was determined the previous stem was loose due to aseptic loosening.The growing distal femur construct was removed and the distal femur revised.All components except the mrh tibia baseplate and stem extender were replaced." rep provided usage sheets from current and previous procedures and reported that no further information will be available.
 
Event or Problem Description
It was reported the patient's left knee was revised.As reported by rep: "patient presented with thigh pain, it was determined the previous stem was loose due to aseptic loosening.The growing distal femur construct was removed and the distal femur revised.All components except the mrh tibia baseplate and stem extender were replaced." rep provided usage sheets from current and previous procedures and reported that no further information will be available.
 
Additional Manufacturer Narrative
Additional information: update to lot code.An event regarding loosening involving a mrs stem was reported.The event was not confirmed.Method & results: -device evaluation and results: not performed as product was not returned.-clinician review: no medical records were received for review with a clinical consultant.-device history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the reported lot.Conclusion: the exact cause of the event could not be determined because insufficient information was provided.Additional information including operative reports, progress notes, x-rays and return of the device are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MRS FEM STEM W/O BODY 13X127MM
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key9197472
Report Number0002249697-2019-03518
Device Sequence Number409643
Product Code KRO
UDI-Device Identifier07613327048377
UDI-Public07613327048377
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K040749
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2019
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/30/2020
Device Catalogue Number64853113
Device Lot Number148636A
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/19/2019
Supplement Date Received by Manufacturer11/13/2019
Initial Report FDA Received Date10/16/2019
Supplement Report FDA Received Date12/11/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age17 YR
-
-