| Catalog Number |
64853113 |
| Medical Device Problem Code |
Osseointegration Problem (3003)
|
| Health Effect - Clinical Codes |
Pain (1994); Injury (2348); Inadequate Osseointegration (2646)
|
| Date of Event |
09/19/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
It was reported the patient's left knee was revised.As reported by rep: "patient presented with thigh pain, it was determined the previous stem was loose due to aseptic loosening.The growing distal femur construct was removed and the distal femur revised.All components except the mrh tibia baseplate and stem extender were replaced." rep provided usage sheets from current and previous procedures and reported that no further information will be available.
|
| |
|
Event or Problem Description
|
|
It was reported the patient's left knee was revised.As reported by rep: "patient presented with thigh pain, it was determined the previous stem was loose due to aseptic loosening.The growing distal femur construct was removed and the distal femur revised.All components except the mrh tibia baseplate and stem extender were replaced." rep provided usage sheets from current and previous procedures and reported that no further information will be available.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: update to lot code.An event regarding loosening involving a mrs stem was reported.The event was not confirmed.Method & results: -device evaluation and results: not performed as product was not returned.-clinician review: no medical records were received for review with a clinical consultant.-device history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the reported lot.Conclusion: the exact cause of the event could not be determined because insufficient information was provided.Additional information including operative reports, progress notes, x-rays and return of the device are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.
|
| |
|