Product event summary: the controller was not returned for evaluation.Review of log file analysis revealed three (3) controller power up with associated motor start events on (b)(6) 2019 at 14:24:02, 14:26:26 and 14:27:18 hours.The data point prior to the loss of power revealed that a battery was connected to power port one (1) with 47% relative state of charge (rsoc) and another battery was connected to power port two (2) with 97% rsoc.No data point was logged after the last controller power up event.No anomalies were observed during the recorded controller power ups.The maximum time controller was without power for 16 minutes and 51 seconds.As a result, the reported event was confirmed.A possible root cause of the loss of power can be attributed to a disconnection of both power sources as described in the event details.An internal investigation was initiated to capture events involving the controller losing power.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|