SMITH & NEPHEW, INC. GII DISHED INS SZ 3-4 9MM; PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY Back to Search Results
Model Number 71420766
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 09/19/2019
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a revision surgery was performed due to infection.Poly was exchanged.Device will not be returned.Xrays not available.
 
Additional Manufacturer Narrative
It was reported that a revision surgery was performed due to infection.Poly was exchanged.The associated genesis ii dished insert, used in treatment, was not returned for evaluation.Thus the product analysis could not be performed.Our investigation including a review of the manufacturing records for the listed batch did not reveal any deviation from the standard manufacturing processes.A review of the complaint history for the listed part revealed no prior complaints for the listed failure mode with the same batch number.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.The devices was sterilized according to sterilization release documentation from quality control.A relationship, if any, between the device and the reported incident could not be corroborated.A review of instructions for use revealed the alleged failure is previously identified in the possible adverse events.Per our risk assessment, possible causes could include but not limited to contamination, patient reaction or post-operative healing issue.No medical documents were received for investigation.Therefore no medical assessment can be performed at this time.Without the return of the actual products involved and no patient medical records available, our investigation of this report is inconclusive.Based on this investigation, the need for corrective action is not indicated.No further investigation is warranted for this complaint; however smith and nephew will continue to monitor for future complaints and investigate as necessary.Should the device or additional information be received, the complaint will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII DISHED INS SZ 3-4 9MM
Common Device Name
PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key9199341
Report Number1020279-2019-03665
Device Sequence Number13565346
Product Code JWH
UDI-Device Identifier03596010209320
UDI-Public03596010209320
Combination Product (Y/N)N
Initial Reporter CountryAS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number71420766
Device Catalogue Number71420766
Device Lot Number18BM12708
Initial Date Received by Manufacturer 09/19/2019
Supplement Date Received by Manufacturer02/04/2020
Initial Report FDA Received Date10/16/2019
Supplement Report FDA Received Date02/07/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-